Optimal Multidetector Row CT Parameters for Evaluations of the Breast: A Phantom and Specimen Study

Authors
Yi, AnnSeo, Bo KyoungCho, Pyong KonPisano, Etta D.Lee, Ki YeolJe, Bo KyungKim, Hee-YoungMin, Byung WookSon, Gil Soo
Issue Date
Jun-2010
Publisher
ELSEVIER SCIENCE INC
Keywords
breast; computed tomography (CT); multidetector row, computed tomography (CT); radiation exposure; phantom imaging, mastectomy
Citation
ACADEMIC RADIOLOGY, v.17, no.6, pp 744 - 751
Pages
8
Indexed
SCIE
SCOPUS
Journal Title
ACADEMIC RADIOLOGY
Volume
17
Number
6
Start Page
744
End Page
751
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/14868
DOI
10.1016/j.acra.2010.02.008
ISSN
1076-6332
1878-4046
Abstract
Rationale and Objectives: The use of breast computed tomography (CT) has been limited by radiation hazard and image quality. The objective of this study was to compare radiation doses and image quality for different 64-channel multidetector row CT (MDCT) parameters, and to potentially provide optimal CT parameters for breast imaging. Materials and Methods: For assessment of radiation doses, CT dose index (CTDI(100)) values were obtained at various x-ray tube voltages (80, 120, 140 kVp) and currents (30, 50, 100, 150, 200 mAs) using a standard CT body dose phantom. To evaluate image quality, four fresh mastectomy specimens were scanned and three radiologists graded images for overall image quality, glandular tissue-fat conspicuity, and Cooper's ligament sharpness. Statistically, linear regression analyses and multiple comparisons were used for investigation of the relationship between radiation dose, image qualities, and CT parameters. Results: CTDI(100) values of <= 6 mGy were obtained at 80 kVp and any mAs, 120 kVp and 30 or 50 mAs, and 140 kVp and 30 or 50 mAs. Image quality at 80 kVp and 200 mAs, 120 kV and 50, 100, 150, or 200 mAs, and 140 kVp and all mAs values tested were significantly superior to those at 80 kVp and 30, 50, 100, or 150 mAs and 120 kV and 30 mAs (P < .05). Conclusions: Based on our results, 80 kVp and 200 mAs, 120 kVp and 50 mAs, 140 kVp and 30 mAs, or 140 kVp and 50 mAs can be used for breast MDCT scanning to reduce radiation dose and preserve image quality and 140 kVp at 30 mAs is the optimal setting.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Breast and Endocrine Surgery > 1. Journal Articles
2. Clinical Science > Department of Colon and Rectal Surgery > 1. Journal Articles
2. Clinical Science > Department of Radiology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Min, Byung Wook photo

Min, Byung Wook
Guro Hospital (Department of Colon and Rectal Surgery, Guro Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE