Detailed Information

Cited 12 time in webofscience Cited 14 time in scopus
Metadata Downloads

Agreement and Reproducibility of Proton Density Fat Fraction Measurements Using Commercial MR Sequences Across Different Platforms A Multivendor, Multi-Institutional Phantom Experiment

Authors
Jang, Jong KeonLee, Seung SooKim, BohyunCho, Eun-SukKim, Yeong JaeByun, Jae HoPark, Beom JinKim, So YeonKim, Jin Hee
Issue Date
Aug-2019
Publisher
Lippincott Williams & Wilkins Ltd.
Keywords
PDFF; commercial MR sequence; CS-MRI; HISTO; reproducibility; phantom study; hepatic steatosis
Citation
Investigative Radiology, v.54, no.8, pp 517 - 523
Pages
7
Indexed
SCI
SCIE
SCOPUS
Journal Title
Investigative Radiology
Volume
54
Number
8
Start Page
517
End Page
523
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/1752
DOI
10.1097/RLI.0000000000000561
ISSN
0020-9996
1536-0210
Abstract
Objectives The aim of this study was to evaluate the agreement and reproducibility of proton density fat fraction (PDFF) measurements using commercial magnetic resonance (MR) sequences across different imagers, vendors, and field strengths via a phantom experiment. Materials and Methods Eleven fat-water emulsion phantoms of varying fat proportions (ie, 0–50 weight%) were constructed. Phantom PDFFs were estimated using commercial chemical shift–based MR imaging sequences with Siemens 1.5 T and 3.0 T, Philips 3.0 T, and GE 1.5 T and 3.0 T imagers, and MR spectroscopic sequences (HISTO) with Siemens 1.5 T and 3.0 T imagers. Agreement among the estimated PDFF values between commercial sequences was evaluated using Bland-Altman analysis. Reproducibility of the PDFF measurements across commercial sequences was evaluated using the reproducibility coefficient. The test-retest repeatability of the PDFF measurements was evaluated using the repeatability coefficient. Results The repeatability coefficient of the PDFF measurements was 0.31% to 1.58% for the absolute PDFF value for commercial sequences. Statistically significant biases in the estimated PDFF were noted in 19 of 21 pairwise comparisons of commercial sequences (range of mean biases, −4.48% to 8.15% for the absolute PDFF value). The reproducibility coefficient of PDFF measurements was 9.0% for the absolute PDFF value over all commercial sequences and 10.6% for the absolute PDFF value over all chemical shift–based MR imaging sequences. Conclusions The measurement of the PDFF is highly repeatable with commercial MR sequences but is not reproducible across different sequences, imager vendors, and field strengths. The use of the same sequence and imager is therefore recommended for the longitudinal follow-up of hepatic steatosis using commercial MR sequences for PDFF measurements.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Radiology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Park, Beom jin photo

Park, Beom jin
Anam Hospital (Department of Radiology, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE