Detailed Information

Cited 13 time in webofscience Cited 15 time in scopus
Metadata Downloads

Pulsatile versus nonpulsatile flow to maintain the equivalent coronary blood flow in the fibrillating heart

Authors
Jung, Jae SeungSon, Ho SungLim, Choon HakSun, Kyung
Issue Date
Nov-2007
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Citation
ASAIO JOURNAL, v.53, no.6, pp 785 - 790
Pages
6
Indexed
SCIE
SCOPUS
Journal Title
ASAIO JOURNAL
Volume
53
Number
6
Start Page
785
End Page
790
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/17655
DOI
10.1097/MAT.0b013e31815b2d00
ISSN
1058-2916
1538-943X
Abstract
How much flow is required by a nonpulsatile pump to match the coronary blood flow equivalent to that of pulsatile pump? A cardiopulmonary bypass circuit from the right atrium to the ascending aorta was constructed in a ventricular fibrillation model using 13 Yorkshire swine. The animals were randomly divided into two groups: CONTROL (pulsatile T-PLS, n = 7) or EXPERIMENTAL (nonpulsatile Biopump, n = 6). The hemodynamic data at mid-LAD level was measured with a flow meter at baseline and every 20 minutes after pump flow initiation. The pump flow was started from 2 L/min in both groups (67 +/- 8 in CONTROL and 70 +/- 9 ml/kg/min in EXPERIMENTAL; p = NS), and the pump flow of the EXPERIMENTAL group was increased to match the coronary flow of the CONTROL group. To maintain mean velocity and flow in the LAD, the EXPERIMENTAL group required significantly higher pump flow at 20, 40, and 60 minutes (84 17 vs. 67 8, 87 +/- 24 vs. 67 +/- 8, 85 +/- 18 vs. 67 +/- 8 ml/kg/min, respectively, p < 0.05). The LAD diameter was substantially smaller in the CONTROL group and the resistance index was significantly lower in the CONTROL group at 80 minutes and 120 minutes after bypass (0.56 +/- 0.26 vs. 0.87 +/- 0.20 and 0.61 +/- 0.23 vs. 0.90 +/- 0.06; p < 0.05). In conclusion, we found that the nonpulsatile pump may require 25%-28% higher pump flow than the pulsatile pump to maintain equivalent coronary blood flow.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Anesthesiology and Pain Medicine > 1. Journal Articles
2. Clinical Science > Department of Thoracic and Cardiovascular Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Choon Hak photo

Lim, Choon Hak
Anam Hospital (Department of Anesthesiology and Pain Medicine, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE