Calibration of Ir-192 HDR source based on standards of absorbed dose to water

Authors
Huh H.D.Kim W.C.Loh J.J.K.Lee S.Shin D.O.Lee S.H.Lim S.Kwon S.Choi J.Kim S.H.
Issue Date
2007
Publisher
Springer Verlag
Keywords
Absorbed dose to water calibration factor; Air kerma strength; Brachytherapy source
Citation
IFMBE Proceedings, v.14, no.1, pp 1877 - 1879
Pages
3
Indexed
SCOPUS
Journal Title
IFMBE Proceedings
Volume
14
Number
1
Start Page
1877
End Page
1879
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/31799
ISSN
1680-0737
Abstract
Brachytherapy sources are usually calibrated in terms of the measured air-kerma strength SK in free space at 1 m distance. The trend of the protocols for the external beam radiotherapy has been changing from an air_kerma base to standards of absorbed dose to water. In this work we tried to specify the source strength of Ir-192 source in terms of the absorbed_dose_to_water instead of an air kerma strength using standards of absorbed dose to water. For this work the MPBP(the Multi Purpose Brachytherapy Phantom) was home-made, which was designed to put both sources and chambers into water at the exact point we want. The reference point was chosen at 5 cm distance along the transverse axis of the source. The computer system of Plato RTP, BPS v13.2, Nucletron, Holland, was used to deliver the dose we want to the reference point. The absorbed_ dose_to_water calibration factor NDW of ionization chambers for brachytherapy sources used must be determined in order to apply standards of absorbed dose to water. The PTW TN 30013 chamber was used and the calibration factor NDW of the chamber was determined using the Monte Carlo Code of EGSnrcCPP. The factor calculated was then used to determine the absorbed dose to water, which calculated an air kerma strength for an Ir-192 source at the reference point. The dose calculated was compared with the dose delivered by Plato RTP. The dose delivered to the reference point by the computer system used was 100 cGy. The calibration factor NDW calculated was 5.28 cGy /nC and was used to calculate the absorbed doses at some distances, which show 2.16% to 0.84% difference from the dose delivered. Air kerma strength calculated showed the difference from -2.7% to +1.7% relative to the one of the source provided by a vendor. This shows the feasibility of using standards of absorbed dose to water to calibrate the brachytherapy source. © International Federation for Medical and Biological Engineering 2007.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Radiation Oncology > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Suk photo

Lee, Suk
Anam Hospital (Department of Radiation Oncology, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE