Detailed Information

Cited 5 time in webofscience Cited 5 time in scopus
Metadata Downloads

Detailed comparison of the da Vinci Xi and S surgical systems for transaxillary thyroidectomy

Authors
Yu, Da YoungChang, Young WooLee, Hye YoonKim, Woo YoungKim, Hoon YubLee, Jae BokSon, Gil Soo
Issue Date
22-Jan-2021
Publisher
Lippincott Williams & Wilkins Ltd.
Keywords
operative procedures; robotic surgical procedures; surgical injuries; thyroid carcinoma; thyroidectomy
Citation
Medicine, v.100, no.3
Indexed
SCIE
SCOPUS
Journal Title
Medicine
Volume
100
Number
3
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/52142
DOI
10.1097/MD.0000000000024370
ISSN
0025-7974
1536-5964
Abstract
Robotic surgical systems have evolved over time. The da Vinci Xi system was developed in 2014 and was expected to solve the shortcomings of the previous S system. Therefore, we conducted this study to compare these 2 systems and identify if the Xi system truly improves surgical outcomes. In this retrospective study, a total of 86 patients with unilateral papillary thyroid carcinoma without central lymph node involvement underwent gasless transaxillary hemithyroidectomy using 2 robotic systems, the da Vinci S and Xi. Forty patients were in the da Vinci S group and 46 patients were in the da Vinci Xi group. All surgeries were performed by 1 surgeon (YWC). All surgery video files were analyzed to compare the duration of each surgical step. The total operation time was significantly shorter in the Xi group than in the S group (153.0 minutes vs 105.7 minutes, P < .01). Time for robot docking was shorter in the Xi group (19.8 minutes vs 10.6 minutes, P < .01), and all procedures performed in the console also required a shorter time in this group. The overall complication rate did not differ significantly (P = .464). The da Vinci Xi system made robotic thyroidectomy easier and faster without increasing the complication rate. It is a safe and valuable system for robotic thyroidectomy.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Anesthesiology and Pain Medicine > 1. Journal Articles
2. Clinical Science > Department of Breast and Endocrine Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Jae Bok photo

Lee, Jae Bok
Guro Hospital (Department of Breast and Endocrine Surgery, Guro Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE