Durability improvement of polymer chamber of pulsatile extracorporeal life support system in terms of mechanical change

  • Choi H.; 
  • Paik S.H.; 
  • Lee K.H.; 
  • Min B.G.; 
  • Won Y.S.
Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

Twin Pulse Life Support, T-PLS™ has received the CE mark (2003) and Korea Food and Drug Administration (KFDA) approval (2004) for short-term application as an Extracorporeal Life Support system (ECLS). T-PLS's original intention was to apply for not only short-term but also long-term application such as Extracorporeal ventricular assist device (VAD). Hence, a long-term durability test was conducted. The 1-year reliability of the systems tested in this study did not meet the STS/ASAIO standard of 80% reliability with 60% confidence for a 1-year mission life. However, without the disposable units, which are only designed to operate for 6 h, the 1-year reliability exceeded the STS/ ASAIO standard of 80% reliability with 60% confidence. In this study, by using the existing analysis methods and analyzing the root cause of the failure used by a numerical analysis. As eliminating or mitigating of the root cause of the failure, we improved the durability of blood chamber and evaluated the performance of the modified system via the hemolysis test. © International Federation for Medical and Biological Engineering 2007.

키워드

Blood chamber; Durability; Extracorporeal life support systems (ECLS); Hemolysis testing; Reliability; Blood; Durability; Numerical analysis; Reliability; Blood chamber; Extracorporeal life support system; Hemolysis testing; Biomedical equipment; polyetherurethan; polymer; polyurethan; polyvinylchloride; analytic method; animal experiment; article; assisted circulation; blood analysis; confidence interval; controlled study; disposable equipment; equipment design; extracorporeal circulation; food and drug administration; Korea; mathematical analysis; medical instrumentation; nonhuman; reliability; standardization; Equipment Design; Equipment Failure; Equipment Failure Analysis; Extracorporeal Membrane Oxygenation; Hemolysis; Humans; Materials Testing; Polymers; Pulsatile Flow
제목
Durability improvement of polymer chamber of pulsatile extracorporeal life support system in terms of mechanical change
저자
Choi H.; Paik S.H.; Lee K.H.; Min B.G.; Won Y.S.
DOI
10.1007/s11517-007-0215-9
발행일
2007-11
유형
Article
저널명
Medical & Biological Engineering & Computing
권
45
호
11
페이지
1127 ~ 1135