나사못으로 고정한 무시멘트형 비구컵의 불충분한 덮임이 접촉면의 안정성에 미치는 영향

Interface Stability on Insufficient Coverage of Non-cemented Acetabular Cup with Screw Fixations

초록

Purpose To analyze the effects of various amount of peripheral defect of the acetabulum in terms of the stability of the acetabular cup with screw fixation. Materials and Methods The porous coated acetabular cup of Trilogy was press-fitted with 3 screw fixations after 2 mm undersized reaming into the distal part of 100 bovine femurs, but with insufficient peripheral coverage of 0%, 5%, 10%, 15% and 20% respectively. The micromotions of the acetabular cups were measured under the axial loads of 50, 100, 150, 200, 250, and 300 Kp, and torques were measured at 1° to 5° of cup rotation under applied rotational force in 10 specimens in each of five groups. Results The micromotion between the cup and bone interface increased proportionally with peripheral coverage insufficiency and the torque values in the five groups increased in proportion to the rotation degree and were inversely proportinal peripheral coverage insufficiency. Conclusion 15% or less coverage insufficiency of the acetabular cup may be acceptable as the average micromotion is less than 150 m and compatible with biologic fixations.

키워드

Porous coated Trilogy acetabular component; Insufficient coverage; Press fit; Screw fixations; Interface stability
제목
나사못으로 고정한 무시멘트형 비구컵의 불충분한 덮임이 접촉면의 안정성에 미치는 영향
제목 (타언어)
Interface Stability on Insufficient Coverage of Non-cemented Acetabular Cup with Screw Fixations
저자
박상원; 우경조; 한승범; 오광준; 주성대
발행일
2002-06
저널명
대한정형외과학회지
권
37
호
3
페이지
437 ~ 442