Comparing the safety of harmonic ACE and ACE plus around the recurrent laryngeal nerve in swine models
- Kim, Hong Kyu;
- Chai, Young Jun;
- Lee, Hye Yoon;
- Kim, Hoon Yub;
- Dionigi, Gianlorenzo
WEB OF SCIENCE
10SCOPUS
9초록
Purpose: Among the various energy-based devices, ultrasonic shears are popular in thyroid surgeries. In this study, we tested the safety of Harmonic ACE and Harmonic ACE+ around the recurrent laryngeal nerve (RLN) in experimental swine models. Methods: Harmonic ACE and Harmonic ACE+ were each tested in 4 piglets. Harmonic ACE and Harmonic ACE+ were activated at a 0- to 5-mm distance from the RLN. The function of the RLN was assessed using continuous electrophysiological monitoring. Results: For Harmonic ACE, there was no adverse EMG event found when activated at 4- and 5-mm distances from the RLN. At a 2- to 3-mm distance, there were 4. adverse EMG events observed. In these 4 cases, adjacent tissue shrinkage occurred after 6 to 15 seconds of activation, and the RLN touched the Harmonic ACE. At a 1-mm distance, there were 2 adverse EMG events found after 25 seconds of activation. For Harmonic ACE+, there was no adverse EMG event observed when activated at 1- and 3-mm distances from the RLN. At a 0-mm distance, 2 adverse EMG events occurred after 6 to 10 seconds of activation. Conclusion: The safe distance of Harmonic ACE and ACE+ was 4 and 1 mm, respectively, in the swine models. Harmonic ACE+ is safer than Harmonic ACE because it did not cause any tissue shrinkage. Surgeons need to understand the characteristics of devices for safe operation.
키워드
- 제목
- Comparing the safety of harmonic ACE and ACE plus around the recurrent laryngeal nerve in swine models
- 저자
- Kim, Hong Kyu; Chai, Young Jun; Lee, Hye Yoon; Kim, Hoon Yub; Dionigi, Gianlorenzo
- 발행일
- 2018-06
- 유형
- Article
- 권
- 94
- 호
- 6
- 페이지
- 285 ~ 290
- 언어
- ENG
- 출판사
- 대한외과학회
- 발행국가
- 대한민국
- 분량
- 6 페이지
- ISSN
- E 2288-6796
P 2288-6575