Self-monitoring of capillary blood glucose is important for controlling diabetes. Recently, a laser lancing device (LMT-1000) that can collect capillary blood without skin puncture was developed. We enrolled 150 patients with type 1 or 2 diabetes mellitus. Blood sampling was performed on the same finger on each hand using the LMT-1000 or a conventional lancet. The primary outcome was correlation between glucose values using the LMT-1000 and that using a lancet. And we compared the pain and satisfaction of the procedures. The capillary blood sampling success rates with the LMT-1000 and lancet were 99.3% and 100%, respectively. There was a positive correlation (r=0.974, <i>P</i><0.001) between mean blood glucose levels in the LMT-1000 (175.8±63.0 mg/dL) and conventional lancet samples (172.5±63.6 mg/dL). LMT-1000 reduced puncture pain by 75.0% and increased satisfaction by 80.0% compared to a lancet. We demonstrated considerable consistency in blood glucose measurements between samples from the LMT-1000 and a lancet, but improved satisfaction and clinically significant pain reduction were observed with the LMT-1000 compared to those with a lancet.

키워드

Blood glucose self-monitoring; Diabetes mellitus; Lasers; Pain; INSULIN; COMPLICATIONS; IMPACT
제목
Comparison of Laser and Conventional Lancing Devices for Blood Glucose Measurement Conformance and Patient Satisfaction in Diabetes Mellitus
저자
Kim, Jung A; Park, Min Jung; Song, Eyun; Roh, Eun; Park, So Young; Lee Da Young; Kim, Jaeyoung; Yu, Ji Hee; Seo, Ji A; Choi, Kyung Mook; Baik, Sei Hyun; Yoo, Hye Jin; Kim, Nan Hee
DOI
10.4093/dmj.2021.0293
발행일
2022-11
저널명
Diabetes and Metabolism Journal
권
46
호
6
페이지
936 ~ 940