Development of an index-matching epoxy-based collimator for reducing multipath artifacts in multimodal intravascular optical coherence tomography

  • Cho, Youngeun; 
  • Han, Jeongmoo; 
  • Kim, Yeon Hoon; 
  • Nam, Hyeong Soo; 
  • Lee, Min Woo; 
  • ... Kim, Jin Won; 
  • 외 2명
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

We developed a novel collimator with an expanded, index-matched cladding for double-clad fiber-based multimodal intravascular optical coherence tomography (OCT) systems. The proposed design mitigates multipath artifacts by reducing cladding-to-core coupling at fiber discontinuities, resulting in an improvement in signal-to-noise ratio. This enhancement effectively suppresses image artifacts and improves overall OCT image quality, thereby enabling more accurate diagnostic assessment in multimodal intravascular imaging.

키워드

Optical coherence tomography; Collimator; Double-clad fiber; Multipath artifacts
제목
Development of an index-matching epoxy-based collimator for reducing multipath artifacts in multimodal intravascular optical coherence tomography
저자
Cho, Youngeun; Han, Jeongmoo; Kim, Yeon Hoon; Nam, Hyeong Soo; Lee, Min Woo; Kim, Young-Jin; Kim, Jin Won; Yo, Hongki
DOI
10.1117/12.3098952
발행일
2026-02
유형
Proceedings Paper
저널명
Proceedings of SPIE - The International Society for Optical Engineering
권
14016