초록
PURPOSE The purpose of this study was to compare the two-dimensional with three-dimensional radiographic measurements of polyethylene liner wear in the Harris-Galante II acetabular cup. MATERIALS AND METHODS We measured the polyethylene liner wear amount and the wear rate of 64 hips with Harris-Galante II acetabular cup by two methods: Livermore's method and modified Devane's method. We evaluated the relationships between the two methods. RESULTS The average total amount of wear and the average wear rate are 0.86±1.01 mm and 0.18±0.20 mm/year using the two-dimensional method and 0.99±1.01 mm and 0.21±0.2 mm/year using three-dimensional method respectively. The amount of wear and the wear rate by the three-dimensional method was larger than those by the two dimensional method. CONCLUSION The amount of polyethylene liner wear can be measured by the two-dimensional method because it is highly correlated with the amount that is measured by the three-dimensional method, since there was no differences between the two methods.
키워드
- 제목
- Harris-Galante Acetabular Cup에서 폴리에틸렌 라이너 마모의 계측 : 이차원적 계측과 삼차원적 계측의 비교
- 제목 (타언어)
- Polyethylene Liner Wear in Harris-Galante Acetabular Cup: Two Dimensional versus Three Dimensional Method
- 저자
- 박상원; 정웅교; 한승범
- 발행일
- 2001-08
- 저널명
- 대한정형외과학회지
- 권
- 36
- 호
- 4
- 페이지
- 373 ~ 376
- 언어
- KOR
- 출판사
- 대한정형외과학회
- 발행국가
- 대한민국
- 분량
- 4 페이지
- ISSN
- E 2005-8918
P 1226-2102