척추성형술 시술에서 PMMA주입에 대한 흉추의 생체역학적 평가

Biomechanical Evaluation of PMMA Injection in Vertebroplasty

초록

In this study, method of PMMA injection is suggested for vertebroplasty in patients with osteoporotic compression fracture. The finite element analysis is used to investigate the vertebroplasty quantitatively. In order to improve previous works with simplified geometry of vertebral body more exact geometry has been constructed from CT image data with 1mm thickness. An ideal method of PMMA delivery, with respect to location and amount of injectate, into vertebral body has been suggested based on evaluation of the insert positions and the insert shapes of injected PMMA. It is shown that vertebral body can be compensated most efficiently when PMMA is highly concentrated on the top-front of trabecular bone of compressed vertebra.

키워드

Vertebroplasty; Osteoporosis; Biomechanical analysis; Finite element analysis; PMMA(Polymethylmethacrylate); Compression fracture
제목
척추성형술 시술에서 PMMA주입에 대한 흉추의 생체역학적 평가
제목 (타언어)
Biomechanical Evaluation of PMMA Injection in Vertebroplasty
저자
이준형; 채수원; 이태수; 서중근; 박정율; 김상돈; 이관행
발행일
2004-02
유형
Article
저널명
의공학회지
권
25
호
1
페이지
27 ~ 32