Position-wise comparison of handheld 6-lead ECG versus the standard 12-lead ECG in patients with arrhythmia: comparative study
- Park, Young Jun;
- Eom, Sujeong;
- Lee, Sang Jun;
- Na, Yeongyeon;
- Lim, Jeonghwa;
- 외 5명
WEB OF SCIENCE
0SCOPUS
0초록
Background: As 12‑lead electrocardiographs (ECGs) require a clinical infrastructure that limits timely access, portable 6‑lead devices may extend diagnostics to community and remote settings. We evaluated the signal equivalence of a handheld 6‑lead ECG (HATIV® P30) versus the standard 12‑lead in an arrhythmia cohort, considering posture and synchrony. Methods: In this prospective single-center study, simultaneous 10-s 12-lead ECGs and time-aligned 10-s segments from 30-s 6-lead recordings were obtained from arrhythmia patients in both supine and sitting positions. A blinded electrophysiologist performed rhythm classification and ECG measurements. Diagnostic accuracy and numerical agreement of key parameters (PR interval, QRS duration, QT/QTc intervals, and amplitudes) were evaluated. Results: A total of 229 paired recordings were analyzed after excluding 6 pairs. The overall diagnostic accuracy of the 6‑lead versus 12‑lead was 99.1% in the supine position (n = 113) and 99.1% in the sitting position (n = 116); one atrial flutter was misclassified as atrial fibrillation in each position. Bland-Altman analyses showed small mean differences (12‑lead minus 6‑lead): PR + 12.1/ + 7.4 ms (supine/sitting), QRS - 6.4/ - 6.0 ms, QT - 10.3/ - 5.3 ms, QTc - 11.5/ - 6.4 ms; heart‑rate difference ≈0.03 bpm. The absolute differences were < 20 ms in approximately ~ 70% for PR and ~ 55-62% for QT/QTc. In an exploratory asynchronous pairing (supine 12‑lead vs sitting 6‑lead; n = 103), accuracy decreased to 97.1% and parameter differences widened, consistent with postural/temporal effects. Conclusions: In patients with arrhythmia, the handheld 6‑lead showed near‑perfect rhythm agreement and small numerical differences versus the 12‑lead under synchronized acquisition in both positions. Asynchronous or posture-mismatched comparisons reduce the agreement, and acquisition conditions should be considered. The 6‑lead may be a practical alternative when the 12‑lead is unavailable in patients with arrhythmia. © 2026. The Author(s).
키워드
- 제목
- Position-wise comparison of handheld 6-lead ECG versus the standard 12-lead ECG in patients with arrhythmia: comparative study
- 저자
- Park, Young Jun; Eom, Sujeong; Lee, Sang Jun; Na, Yeongyeon; Lim, Jeonghwa; Park, Minje; Ahn, Hyun Jin; Chang, Mineok; Joo, Sunghoon; Ahn, Min-Soo
- 발행일
- 2026-06
- 유형
- Article
- 권
- 26
- 호
- 1
- 언어
- ENG
- 출판사
- BioMed Central
- 발행국가
- 영국
- ISSN
- P 1471-2261