상세 보기
Tenelectrodes: a new stimulator for inching technique in the diagnosis of carpal tunnel syndrome
- Kang, YK;
- Kim, DH;
- Lee, SH;
- Hwang, M;
- Han, MS
WEB OF SCIENCE
7SCOPUS
7초록
This study was designed to evaluate the usefulness of a new multielectrode stimulator, TenElectrodes, in the diagnosis and localization of the compression site in the wrists of carpal tunnel syndrome (CTS) patients. Antidromic inching technique (IT) of the median nerve at the wrist was performed with the TenElectrodes, on 46 controls and 21 CTS patients. In controls, mean conduction delay per centimeter (CD/cm) was 0.21 milliseconds (ms), and maximal CD/crn was 0.27 ms in the segment 3 to 4 centimeters distal to the distal wrist crease. The abnormal cut-off value, calculated as the maximal CD/cm + 2SD, was 0.45 ms. In the CTS group, the maximal CD/cm was 0.56 ms in the segment 2 to 3 centimeters distal to the distal wrist crease, and the CD/crn values in all segments between the distal wrist crease and 4 cm distal to the distal wrist crease were greater than 0.45 ms. Antidromic IT using TenElectrodes may be an easy, fast and accurate method as the electrodes of the stimulator are aligned at 1-cm intervals and are adjustable to the wrist contour by springs.
키워드
- 제목
- Tenelectrodes: a new stimulator for inching technique in the diagnosis of carpal tunnel syndrome
- 저자
- Kang, YK; Kim, DH; Lee, SH; Hwang, M; Han, MS
- 발행일
- 2003-06-30
- 유형
- Article
- 권
- 44
- 호
- 3
- 페이지
- 479 ~ 484
- 언어
- ENG
- 출판사
- 연세대학교의과대학
- 발행국가
- 대한민국
- 분량
- 6 페이지
- ISSN
- E 1976-2437
P 0513-5796