Performance of Norudia Glycated Albumin Assay on Multiple Analytical Platforms and Comparison to Lucica Assay
- Choe, Wonho;
- Kim, Sollip;
- Chang, Jeonghyun;
- Park, Hyunwoong;
- Kim, Ha-Nui;
- 외 1명
WEB OF SCIENCE
2SCOPUS
2초록
Background: The Norudia glycated albumin (GA) assay was evaluated for analytical performance and assay applicability on multiple analytical platforms. Methods: The evaluation included precision, linearity, reference interval, and comparison with Lucica GA assay. A multicenter study was conducted to compare the results of Norudia GA assay on five kinds of widely used automated clinical chemistry analyzers. Results: Within-laboratory imprecisions for GA% presented 1.3 - 3.3% and 0.8 - 2.6% for low- and high-level control materials, respectively, on different analyzers. GA assay was linear from 20.0 to 680.0 mu mol/L of GA. The claimed reference range (12 - 16 GA%) was verified. Norudia GA showed a good GA% correlation with Lucica GA (correlation coefficient 0.999). GA% from each analyzer showed good correlation with the consensus mean of the results of five analyzers (correlation coefficient 0.997 - 0.999). Conclusions: The Norudia GA assay can successfully be implemented in all the tested platforms, with good GA% correlation.
키워드
- 제목
- Performance of Norudia Glycated Albumin Assay on Multiple Analytical Platforms and Comparison to Lucica Assay
- 저자
- Choe, Wonho; Kim, Sollip; Chang, Jeonghyun; Park, Hyunwoong; Kim, Ha-Nui; Yoo, Soo Jin
- 발행일
- 2020-11
- 유형
- Article
- 권
- 66
- 호
- 11
- 페이지
- 2383 ~ 2387
- 언어
- ENG
- 출판사
- Clinical Laboratory Publications
- 발행국가
- 독일
- 분량
- 5 페이지
- ISSN
- P 1433-6510