The Effect of Repeated Use of the Disposable Blade in Automated Microkeratome on the Corneal Flap Dimensions and Cut Quality

초록

Purpose: To evaluate changes in the blade, flap dimensions, and stromal bed with repeated use of disposable blade in LASIK (laser in situ keratomileusis) surgery. Methods: A keratome was used to make a corneal flap in pig cadaver eyes, with the blades being reused up to five times. The thickness and diameter of each corneal flap was measured. Scanning electron microscopy of the stromal beds and blade cutting edges was performed to determine the effect that repeated use of the blade had on the stromal bed, and on the cutting edge of the blade. Results: The thickness/diameter of flap was 169.73±7.55 μm/8.60±0.23 mm after single use and 142.93±13.10 μm/8.72±0.20 mm after being used five times, with both demensions decreasing as the blade was used more. The edge of the blade became blunt and irregular and the cutting surface rough and irregular with continual use. Conclusions: During LASIK surgery, reused blades became blunt, causing a thin flap, rough cutting face, and complications such as irregular astigmatism clinically.

키워드

Corneal flap dimensions; Cut quality; LASIK; Microkeratome; Repeated blade
제목
미세각막절개도에서 일회용 칼날의 반복 사용이 칼날과 각막 절삭면에 미치는 영향
제목 (타언어)
The Effect of Repeated Use of the Disposable Blade in Automated Microkeratome on the Corneal Flap Dimensions and Cut Quality
저자
오청훈; 송종석; 김효명
발행일
2004-06
저널명
대한안과학회지
권
45
호
6
페이지
1024 ~ 1029