Spontaneous Late Dissociation of the Tibial Insert After High-Flex Posterior-Stabilized Genesis II Total Knee Arthroplasty

Citations

WEB OF SCIENCE

7
Citations

SCOPUS

9

초록

We report a rare complication of spontaneous late dissociation of the tibial insert 2 years after total knee arthroplasty using a high-flex posterior-stabilized Genesis II prosthesis (Smith & Nephew, Memphis, Tenn). It appears that 2 factors may have contributed to dissociation, namely, incomplete seating of the insert and the design of the prosthesis, which involves a shallow anterior tab snap-fit locking mechanism and thin dovetail lips.

키워드

dissociation; Genesis II; polyethylene insert; total knee arthroplasty; POLYETHYLENE; COMPONENT
제목
Spontaneous Late Dissociation of the Tibial Insert After High-Flex Posterior-Stabilized Genesis II Total Knee Arthroplasty
저자
Lee, Dae-Hee; Lee, Tze Gin; Park, Sung-Joon; Han, Seung-Beom
DOI
10.1016/j.arth.2012.03.056
발행일
2013-02
유형
Article
저널명
Journal of Arthroplasty
권
28
호
2
페이지
374.e13 ~ 374.e16