상세 보기
Comparison of Biotin Interference in Second- and Third-Generation Roche Free Thyroxine Immunoassays
- Choi, Jungim;
- Yun, Seung Gyu
WEB OF SCIENCE
10SCOPUS
11초록
Dear Editor, Immunoassays (IAs) based on streptavidin-biotin binding are widely used in clinical laboratory testing owing to the high affinity and stable interaction between streptavidin and biotin (vitamin B7) and the development of various biotinylation methods [1]. Biotin is a water-soluble B-complex vitamin and a coenzyme responsible for carboxyl transfer in essential carboxylases. Circulating serum concentrations of biotin in the general population typically range from 0.1 ng/mL to 0.8 ng/mL [2]. Biotin is rapidly absorbed, reaches peak plasma concentrations within 1–2 hours, and has an effective serum half-life of 15 hours [3]. Oral administration of biotin doses of 10 mg resulted in peak plasma concentrations ranging from 53 ng/mL to 141 ng/mL [3, 4]. Considering a maximum dosage of 10 mg once a day (q.d.; 10 mg is >300-fold the adequate daily intake) via over the counter (OTC) biotin products, the serum biotin concentration would drop below the in vitro interference threshold of ≤30 ng/mL after eight hours [3]. A high biotin concentration in the blood can interfere with IAs based on streptavidin-biotin binding and is known to cause false high results in competitive IAs and false low results in sandwich IAs [1]. Although the recommended daily intake of biotin is low (30 μg/day), the use of high-dose biotin supplements (up to 10 mg), which are available OTC, has increased in recent years due to unfounded claims that biotin exerts beneficial effects on hair, nails, and skin [5]. According to one survey, 7.7% (95% confidence interval [CI], 6.6–8.9%) of an outpatient population (N=1,944) indicated biotin use; measuring biotin in plasma samples from emergency department patients (N=1,442) showed that 7.4% (95% CI, 6.2–8.9%) had a concentration at or above the lowest known threshold (10 ng/mL) for biotin interference in a Roche Diagnostics IA (Indianapolis, IN, USA) [6].
- 제목
- Comparison of Biotin Interference in Second- and Third-Generation Roche Free Thyroxine Immunoassays
- 저자
- Choi, Jungim; Yun, Seung Gyu
- 발행일
- 2020-05
- 유형
- Letter
- 권
- 40
- 호
- 3
- 페이지
- 274 ~ 276
- 언어
- ENG
- 출판사
- 대한진단검사의학회
- 발행국가
- 대한민국
- 분량
- 3 페이지
- ISSN
- E 2234-3814
P 2234-3806