Comparison of Filtered Back Projection, Hybrid Iterative Reconstruction, Model-Based Iterative Reconstruction, and Virtual Monoenergetic Reconstruction Images at Both Low- and Standard-Dose Settings in Measurement of Emphysema Volume and Airway Wall Thickness: A CT Phantom Study

Citations

WEB OF SCIENCE

19
Citations

SCOPUS

24

초록

Objective: To evaluate the accuracy of emphysema volume (EV) and airway measurements (AMs) produced by various iterative reconstruction (IR) algorithms and virtual monoenergetic images (VME) at both low- and standard-dose settings. Materials and Methods: Computed tomography (CT) images were obtained on phantom at both low- (30 mAs at 120 kVp) and standard-doses (100 mAs at 120 kVp). Each CT scan was reconstructed using filtered back projection, hybrid IR (iDose(4); Philips Healthcare), model-based IR (IMR-R1, IMR-ST1, IMR-SP1; Philips Healthcare), and VME at 70 keV (VME70). The EV of each air column and wall area percentage (WA%) of each airway tube were measured in all algorithms. Absolute percentage measurement errors of EV (APE(vol)) and AM (APE(WA%)) were then calculated. Results: Emphysema volume was most accurately measured in IMR-R1 (APE(vol) in low-dose, 0.053 +/- 0.002; APE(vol) in standard-dose, 0.047 +/- 0.003; all p < 0.001) and AM was the most accurate in IMR-SP1 on both low- and standard-doses CT (APE(WA%) in low-dose, 0.067 +/- 0.002; APE(WA%) in standard-dose, 0.06 +/- 0.003; all p < 0.001). There were no significant differences in the APE(vol) of IMR-R1 between low- and standard-doses (all p 0.05). VME70 showed a significantly higher APE(vol) than iDose(4), IMR-R1, and IMR-ST1 (all p < 0.004). VME70 also showed a significantly higher APE(WA%) compared with the other algorithms (all p < 0.001). Conclusion: IMR was the most accurate technique for measurement of both EV and airway wall thickness. However, VME70 did not show a significantly better accuracy compared with other algorithms.

키워드

Computed tomography; Model-based iterative reconstruction; Virtual monoenergetic image; Emphysema volume; Airway wall thickness; OBSTRUCTIVE PULMONARY-DISEASE; DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; CHEST CT; LUNG; IMPACT; QUALITY; QUANTIFICATION; REDUCTION
제목
Comparison of Filtered Back Projection, Hybrid Iterative Reconstruction, Model-Based Iterative Reconstruction, and Virtual Monoenergetic Reconstruction Images at Both Low- and Standard-Dose Settings in Measurement of Emphysema Volume and Airway Wall Thickness: A CT Phantom Study
저자
Kim, Cherry; Lee, Ki Yeol; Shin, Chol; Kang, Eun-Young; Oh, Yu-Whan; Ha, Moin; Ko, Chang Sub; Cha, Jaehyung
DOI
10.3348/kjr.2018.19.4.809
발행일
2018-07
유형
Article
저널명
Korean Journal of Radiology
권
19
호
4
페이지
809 ~ 817