Preliminary dose calibration results using a TENOMAG polymer gel dosimeter and optical CT (VISTA™)

Citations

SCOPUS

0

초록

We present a preliminary result of dose calibration for 3D dose evaluation using gel dosimetry in radiotherapy. In this work, we measured the percentage depth dose (PDD) and dose linearity using a normoxic polymer gel dosimeter (TENOMAG) and a commercial cone-beam optical CT scanner (VISTA™, Modus Medical Devices, Inc., London, ON, Canada). We confirmed that the PDD measured for 1 and 2 Gy had a build up region at 0.4 cm depth. However, the PDD measured for 3 and 4 Gy had not a build up region. In all cases, the PDD has not greatly decreased with depth above 4 cm depth. Moreover we confirmed that the optical density proportionally increased with dose at 1, 1.5 and 2 cm depth, while the response has not greatly increased with dose for other depths. In this study, we confirmed that we will be able to perform the dose calibration using optical density obtained from a TENOMAG gel data scanned with an optical CT scanner. © 2010 IOP Publishing Ltd.

키워드

Biomedical equipment; Calibration; Density (optical); Density measurement (optical); Dosimeters; Dosimetry; Functional polymers; Scanning; Build-up region; Depth dose; Dose linearity; Gel dosimetry; Medical Devices; Normoxic polymer gel; Optical CT; Polymer gel dosimeter; Computerized tomography
제목
Preliminary dose calibration results using a TENOMAG polymer gel dosimeter and optical CT (VISTA™)
저자
Lee S.; Shim J.B.; Chang K.H.; Cao Y.; Yi J.; Park J.; Cho S.J.; Lee S.H.; Huh H.; Cho K.H.; Min C.K.; Yang D.S.; Park Y.J.; Yoon W.S.; Kim C.Y.
DOI
10.1088/1742-6596/250/1/012042
발행일
2010
유형
Conference Paper
저널명
Journal of Physics: Conference Series
권
250
페이지
189 ~ 192