Applicability of interrupter resistance measurements for evaluation of exercise-induced bronchoconstriction in children

Citations

WEB OF SCIENCE

8
Citations

SCOPUS

8

초록

The interrupter technique is a noninvasive method for measuring air-flow resistance during tidal breathing. This method requires minimal cooperation, and is therefore promising for use in uncooperative children. The aim of this study was to evaluate applicability interrupter resistance (Rint) measurements in the assessment of exercise-induced bronchoconstriction (EIB). Fifty children aged 5-12 years with mild to moderate asthma were tested by exercise challenge, consisting of free outdoor running for 6 min at 80-90% of maximal predicted heart rate for age. Rint, forced expiratory volume in 1 sec (FEV1), and peak expiratory flow (PEF) were measured before and 10 min after exercise. EIB was defined as a fall of 10% or more in FEV1 after exercise. The repeatability of Flint was assessed, and its response to exercise challenge was compared with current standardized methods. The mean intermeasurement coefficient of variation was 4.6%(SD, +/- 3.0%), and the repeatability coefficient was 0.056 kPa/l/sec. Eighteen(36%)of the 50 children had EIB after exercise challenge test. The area under the receiver-operating characteristic (ROC) curve was 0.953 (95% confidence interval, 0.853-0.992; P< 0.001), and the optimal Flint cutoff value was 15.2%, producing a sensitivity of 88.9% and a specificity of 96.9%. The positive and negative predictive values were 94.1% and 93.9%, respectively. The kappa value between FEV1 and Rint was 0.83. The repeatability of Flint measurements was good, and the results of exercise challenge tests using Flint measurements have excellent agreement with the current standardized methods to detect EIB. Considering that only minimal comprehension and coordination are needed without forced breathing technique, the Rint measurement can provide a useful alternative for assessment of EIB in children unable to perform reliable spirometry.

키워드

exercise-induced bronchoconstriction; interrupter resistance; asthma; children; spirometry; cutoff value; BRONCHIAL RESPONSIVENESS; ASTHMA; FLOW; BRONCHODILATOR; METHACHOLINE; CHALLENGE; RESPONSES; ACCURACY; HEALTHY
제목
Applicability of interrupter resistance measurements for evaluation of exercise-induced bronchoconstriction in children
저자
Song, DJ; Woo, CH; Kang, H; Kim, HJ; Choung, JT
DOI
10.1002/ppul.20340
발행일
2006-03
유형
Article
저널명
Pediatric Pulmonology
권
41
호
3
페이지
228 ~ 233