An electromagnetically-actuated polymer micro-pen for picoliter biological assay patterning

Citations

WEB OF SCIENCE

1

초록

This paper reports a polymer micro-pen with an integrated microchannel and a sample reservoir, which can be actuated by Lorentz force induced on an integrated metal actuator for biological assay patterning. The length, width and thickness of the micro-pen are 500 mu m, 120 mu m and 6 mu m, respectively. The fabricated micro-pen is successfully deflected by electromagnetic force from external permanent magnets which generates a magnetic field of 0.2T. Spring constant of the micropen is analytically estimated as 1.9N/m and the measured resonance frequency is about 7.3 kHz. The microchannel formed on this polymer micro-pen is 15 mu m wide and the sample reservoir is 500 mu m thick in the area of 300 mu m x 300 mu m. Red ink is pulled up to the end point of the microchannel from the reservoir by capillary force. A red-ink-dot in diameter of 11 mu m has been placed on paper by the fabricated micro-pen. This demonstrates a sub-picoliter patterning of biological assay is possible.

키워드

micro-pen; electromagnetic actuator; polymer; SU-8; microchannel; CANTILEVER
제목
An electromagnetically-actuated polymer micro-pen for picoliter biological assay patterning
저자
Im, Maesoon; Cho, Il-Joo; Yun, Kwang-Seok; Yoon, Euisik
DOI
10.1109/SENSOR.2005.1497390
발행일
2005-08
유형
Proceedings Paper
저널명
Transducers '05, Digest of Technical Papers, Vols 1 and 2
페이지
1588 ~ 1591