상세 보기
Comparison of Nd:YAG Laser Capsulotomy Rates Between Refractive Segmented Multifocal and Multifocal Toric Intraocular Lenses
- Kim, Jung Wan;
- Eom, Youngsub;
- Yoon, Eun Gyu;
- Choi, Young;
- Song, Jong Suk;
- ... Kim, Hyo Myung;
- 외 2명
WEB OF SCIENCE
12SCOPUS
12초록
Purpose: To investigate the early incidence of neodymium-doped yttrium aluminum garnet (Nd:YAG) laser capsulotomy according to intraocular lens (IOL) type (nontoric vs toric) and surgical techniques (femtosecond laser–assisted cataract surgery vs conventional phacoemulsification) in eyes with refractive multifocal IOLs. Design: Retrospective case-control study. Methods: Nine hundred thirteen eyes from 483 patients implanted with Lentis Mplus LS-313 MF20 (767 eyes) or Lentis Mplus Toric LU-313 MF20T (146 eyes) IOLs (Oculentis GmbH, Berlin, Germany) were enrolled. We compared the incidence of Nd:YAG laser capsulotomy between the nontoric and toric groups. In addition, the incidence of Nd:YAG laser capsulotomy was also evaluated according to the surgical technique used. Results: The overall incidence of Nd:YAG laser capsulotomy was 10.2% (93/913 eyes). The Nd:YAG laser capsulotomy rate was significantly higher in the toric group (24/146; 16.4%) than in the nontoric group (69/767; 9.0%; P = .007). Of the 913 enrolled eyes, 448 eyes (49.1%) underwent femtosecond laser–assisted cataract surgery and 465 eyes (50.9%) underwent conventional phacoemulsification cataract surgery. There was no significant difference in the incidence of Nd:YAG laser capsulotomy between eyes with femtosecond laser–assisted cataract surgery and eyes with conventional phacoemulsification cataract surgery. Conclusion: Patients with refractive multifocal toric IOLs had higher early incidence rates of Nd:YAG laser capsulotomy when compared to those with refractive multifocal nontoric IOLs. Furthermore, femtosecond laser–assisted cataract surgery could not reduce the early incidence of Nd:YAG laser capsulotomy in this study. © 2020 Elsevier Inc.
키워드
- 제목
- Comparison of Nd:YAG Laser Capsulotomy Rates Between Refractive Segmented Multifocal and Multifocal Toric Intraocular Lenses
- 저자
- Kim, Jung Wan; Eom, Youngsub; Yoon, Eun Gyu; Choi, Young; Song, Jong Suk; Jeong, Ji Won; Park, Seh Kwang; Kim, Hyo Myung
- 발행일
- 2021-02
- 유형
- Article
- 권
- 222
- 페이지
- 359 ~ 367
- 언어
- ENG
- 출판사
- Elsevier BV
- 발행국가
- 미국
- 분량
- 9 페이지
- ISSN
- E 1879-1891
P 0002-9394