Measurement of serum anti-Mullerian hormone by revised Gen II or automated assay: Reproducibility under various blood/serum storage conditions

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

2

초록

Objective We investigated the agreement between anti-Müllerian hormone (AMH) levels measured with revised Gen II (rev-Gen II) and automated AMH (Access) assays and evaluated the reproducibility of each method under various blood/serum storage conditions. Methods AMH levels in blood samples from 74 volunteers were measured by rev-Gen II and Access assays under various conditions: immediate serum separation and AMH measurement (fresh control); serum stored at –20 ℃ and AMH measured after 48 hours, 1 week, and 2 years; serum stored at 0 to 4 ℃ and AMH measured after 48 hours and 1 week; and blood kept at room temperature and delayed serum separation after 48 hours and 1 week, with immediate AMH measurement. Results In fresh controls, all rev-Gen II-AMH values were higher than comparable Access-AMH values (difference, 8.3% to 19.7%). AMH levels measured with the two methods were strongly correlated for all sample conditions (r=0.977 to 0.995, all p<0.001). For sera stored at –20 ℃ or 0 to 4 ℃ for 48 hours, Access-AMH values were comparable to control measurements, but rev-Gen II-AMH values were significantly lower. AMH levels in sera stored at –20 ℃ or 0 to 4 ℃ for 1 week were significantly lower than in fresh controls, irrespective of method. Across methods, long-term storage at –20 ℃ for 2 years yielded AMH measurements significantly higher than control values. When serum separation was delayed, rev-Gen II-AMH values were significantly lower than control measurements, but Access-AMH values varied. Conclusion The rev-Gen II and Access-AMH assays showed varying reproducibility across blood/serum storage conditions, but automated Access yielded superior stability to rev-Gen II.

키워드

Anti-Mullerian Hormone; Automation; Biological assay; Blood; Reproducibility of results; Serum; MLLERIAN HORMONE; IMMUNOASSAYS; PERFORMANCE; AGREEMENT; STABILITY; QUALITY; BLOOD
제목
Measurement of serum anti-Mullerian hormone by revised Gen II or automated assay: Reproducibility under various blood/serum storage conditions
저자
Lee, Joong Yeup; Kim, Chung Hyon; Choe, Seung-Ah; Seo, Soyeon; Kim, Seok Hyun
DOI
10.5653/cerm.2022.05687
발행일
2023-06
유형
Article
저널명
Clinical and Experimental Reproductive Medicine
권
50
호
2
페이지
107 ~ 116