Balancing efficiency and diagnostic fidelity in SEP monitoring

Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

A recent study introduced a patient-specific algorithm designed to reduce the acquisition time required for obtaining somatosensory evoked potentials during spinal surgery. While the approach is promising, its reliance on amplitude and latency thresholds may overlook subtle waveform features that are crucial in high-risk patients. Broader validation, integration of waveform morphology, and cautious application in clinically compromised populations are warranted. Optimizing intraoperative neurophysiological monitoring requires not only speed but also diagnostic fidelity.

제목
Balancing efficiency and diagnostic fidelity in SEP monitoring
저자
Choi, Jongsuk
DOI
10.1007/s10877-025-01315-w
발행일
2025-12
유형
Letter; Early Access
저널명
Journal of Clinical Monitoring and Computing
권
39
호
6
페이지
1319 ~ 1320