Comparison of oropharyngeal leak pressure and clinical performance of LMA ProSeal and i-gel (R) in adults: Meta-analysis and systematic review

  • Shin, Hye Won; 
  • Yoo, Hae Na; 
  • Bae, Go Eun; 
  • Chang, Jun Chul; 
  • Park, Min Kyung; 
  • 외 3명
Citations

WEB OF SCIENCE

26
Citations

SCOPUS

29

초록

Background A meta-analysis and systematic review of randomized controlled trials to compare the oropharyngeal leak pressure (OLP) and clinical performance of LMA ProSeal™ (Teleflex® Inc., Wayne, PA, USA) and i-gel® (Intersurgical Ltd, Wokingham, UK) in adults undergoing general anesthesia. Methods Searches of MEDLINE®, EMBASE®, CENTRAL, KoreaMed and Google Scholar® were performed. The primary objective was to compare OLP; secondary objectives included comparison of clinical performance and complications. Results Fourteen RCTs were included. OLP was significantly higher with LMA ProSeal™ than with i-gel® (mean difference [MD] −2.95 cmH2O; 95% confidence interval [CI] −4.30, −1.60). The i-gel® had shorter device insertion time (MD −3.01 s; 95% CI −5.80, −0.21), and lower incidences of blood on device after removal (risk ratio [RR] 0.32; 95% CI 0.18, 0.56) and sore throat (RR 0.56; 95% CI 0.35, 0.89) than LMA ProSeal™. Conclusion LMA ProSeal™ provides superior airway sealing compared to i-gel®.

키워드

Airway sealing; equipment; i-gel (R); laryngeal mask airway proseal; leak; meta-analysis; LARYNGEAL MASK AIRWAY; I-GEL; SUPRAGLOTTIC AIRWAY; ASPIRATION; ANESTHESIA
제목
Comparison of oropharyngeal leak pressure and clinical performance of LMA ProSeal and i-gel (R) in adults: Meta-analysis and systematic review
저자
Shin, Hye Won; Yoo, Hae Na; Bae, Go Eun; Chang, Jun Chul; Park, Min Kyung; You, Hae Seun; Kim, Hyun Jung; Ahn, Hyung Sik
DOI
10.1177/0300060515607386
발행일
2016-06
유형
Review
저널명
Journal of International Medical Research
권
44
호
3
페이지
405 ~ 418