To evaluate the short- and mid-term results and complications ensuing the transcatheter closure of patent ductus arteriosus (PDA). Between October 1999 and December 2005, 117 patients (34 males and 83 females) underwent attempted percutaneous closure of PDA with a minimum diameter of more than 3 mm. Follow-up evaluations were conducted at 1 day and 1, 3, 6, 12 months after the performance of the transcatheter closure. The median age of patients at catheterization was 11 yr (range, 0.6 to 68 yr), median weight was 30 kg (range, 6 to 74 kg), and the median diameter of PDA was 4 mm (range, 3 to 8 mm). This procedure was conducted successfully in 114 patients (97.4%), using different devices. Major complications were detected in 4 patients (3.4%); significant hemolysis (2), infective endocarditis (1), failed procedure due to embolization (1). Minor complications occurred in 6 patients (5.1 %); mild narrowing of the descending aorta (2) and mild encroachment on the origin of the left pulmonary artery (4). Although the transcatheter closure of PDA may be considered to be effective, several complications, including hemolysis, embolization, infective endocarditis, and the narrowing of adjacent vessels may occur in certain cases.

키워드

patent ductus arteriosus; catheterization; complications; COOK DETACHABLE COILS; CATHETER CLOSURE; GIANTURCO COILS; OCCLUDER; OCCLUSION; EXPERIENCE; DEVICE; IMMEDIATE; INFANT; TRIAL
제목
Complications after transcatheter closure of patent ductus arteriosus
저자
Jang, Gi Young; Son, Chang Sung; Lee, Joo Won; Lee, Jae Young; Kim, Soo Jin
DOI
10.3346/jkms.2007.22.3.484
발행일
2007-06
유형
Article
저널명
Journal of Korean Medical Science
권
22
호
3
페이지
484 ~ 490