A Multimodal Multichannel Neural Activity Readout IC with 0.7 mu W/Channel Ca2+-Probe-Based Fluorescence Recording and Electrical Recording
- Lee, Taeju;
- Park, Jee-Ho;
- Cha, Ji-Hyoung;
- Chou, Namsun;
- Jang, Doojin;
- ... Cho, Il-Joo;
- 외 3명
WEB OF SCIENCE
9초록
This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR) of neural activity for the cell-type-specific study of heterogeneous neuronal cell populations. The time-based FR circuit senses Ca2+ concentration using Ca(2+)probes while the ER circuit acquires action potentials (APs) and local field potentials (LFPs). The IC is fabricated in 0.18 mu m CMOS, The FR circuit achieves a recording range of 81dB (75pA to 860nA) and consumes the power of 0.7 mu W/Ch. The ER circuit achieves the input-referred noise (IRN) of 2.7 mu V-rms over the bandwidth (BW) of 10kHz, while consuming the power of 4.9 mu W/Ch. The in-vitro measurement is performed for recording Ca2+ concentration and electrical neural signals.
키워드
- 제목
- A Multimodal Multichannel Neural Activity Readout IC with 0.7 mu W/Channel Ca2+-Probe-Based Fluorescence Recording and Electrical Recording
- 저자
- Lee, Taeju; Park, Jee-Ho; Cha, Ji-Hyoung; Chou, Namsun; Jang, Doojin; Kim, Ji-Hoon; Cho, Il-Joo; Kim, Seong-Jin; Je, Minkyu
- 발행일
- 2019-06
- 유형
- Proceedings Paper
- 저널명
- 2019 SYMPOSIUM ON VLSI CIRCUITS
- 페이지
- C290 ~ C291
- 언어
- ENG
- 출판사
- IEEE
- 발행국가
- 미국