A Multimodal Multichannel Neural Activity Readout IC with 0.7 mu W/Channel Ca2+-Probe-Based Fluorescence Recording and Electrical Recording

  • Lee, Taeju; 
  • Park, Jee-Ho; 
  • Cha, Ji-Hyoung; 
  • Chou, Namsun; 
  • Jang, Doojin; 
  • ... Cho, Il-Joo; 
  • 외 3명
Citations

WEB OF SCIENCE

9

초록

This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR) of neural activity for the cell-type-specific study of heterogeneous neuronal cell populations. The time-based FR circuit senses Ca2+ concentration using Ca(2+)probes while the ER circuit acquires action potentials (APs) and local field potentials (LFPs). The IC is fabricated in 0.18 mu m CMOS, The FR circuit achieves a recording range of 81dB (75pA to 860nA) and consumes the power of 0.7 mu W/Ch. The ER circuit achieves the input-referred noise (IRN) of 2.7 mu V-rms over the bandwidth (BW) of 10kHz, while consuming the power of 4.9 mu W/Ch. The in-vitro measurement is performed for recording Ca2+ concentration and electrical neural signals.

키워드

Neural activity readout IC; fluorescence recording; calcium ion sensing; electrical recording; time-based operation
제목
A Multimodal Multichannel Neural Activity Readout IC with 0.7 mu W/Channel Ca2+-Probe-Based Fluorescence Recording and Electrical Recording
저자
Lee, Taeju; Park, Jee-Ho; Cha, Ji-Hyoung; Chou, Namsun; Jang, Doojin; Kim, Ji-Hoon; Cho, Il-Joo; Kim, Seong-Jin; Je, Minkyu
발행일
2019-06
유형
Proceedings Paper
저널명
2019 SYMPOSIUM ON VLSI CIRCUITS
페이지
C290 ~ C291