Absolute CD4(+) Cell Count Using a Plastic Microchip and a Microscopic Cell Counter

Citations

WEB OF SCIENCE

11
Citations

SCOPUS

11

초록

We have designed and evaluated the performance of a simple, rapid, and affordable method for counting CD4(+) T-cells with the use of plastic microchips. This new system is an adaptation of a "no-lyse, no-wash," volumetric single platform assay, and absolute CD4(+) counts are determined with the use of a microscopic scanning cell counter. To assess the CD4(+) count test precision and linearity of the system, measured CD4(+) counts were compared with two other reference assays (single and dual platform flow cytometry) with the use of 123 clinical samples including samples obtained from 35 HIV-infected patients, and artificially diluted samples. A correlation between the results from the use of the new method and from the use of the two other reference assays was r = 0.98 for the clinical samples. A dilution test of the new method demonstrated a linearity of r >= 0.99, with coefficients of variation <= 7.6% for all concentration levels. Our findings suggest that the new CD4(+) counting device can be potentially be applied for other diagnostic procedures that measure quantities of characteristic antigens or other materials on cells. (C) 2009 Clinical Cytometry Society

키워드

antigens; CD4; CD4 lymphocyte count; flow cytometry; microchip; absolute CD4 count; microscopic cell counter; PLATFORM IMAGE CYTOMETER; VOLUMETRIC CAPILLARY CYTOMETRY; HUMAN-IMMUNODEFICIENCY-VIRUS; RESOURCE-POOR SETTINGS; T-LYMPHOCYTES; HIV; ENUMERATION; BLOOD; QUANTITATION; NUMBERS
제목
Absolute CD4(+) Cell Count Using a Plastic Microchip and a Microscopic Cell Counter
저자
Bae, Sook Young; Park, Hae Chul; Oh, Jun Seo; Yoon, Soo-Young; Park, Dae Won; Choi, In Keun; Kim, Hae Joong; Oh, Jong Hyun; Hur, Dae Sung; Chung, Chanil; Chang, Jun Keun; Robinson, J. Paul; Lim, Chae Seung
DOI
10.1002/cyto.b.20470
발행일
2009-09
유형
Article
저널명
Cytometry Part B - Clinical Cytometry
권
76B
호
5
페이지
345 ~ 353