Evaluation of fully automated commercial software for Agatston calcium scoring on non-ECG-gated low-dose chest CT with different slice thickness

Citations

WEB OF SCIENCE

19
Citations

SCOPUS

22

초록

Objectives To evaluate commercial deep learning–based software for fully automated coronary artery calcium (CAC) scoring on non-electrocardiogram (ECG)-gated low-dose CT (LDCT) with different slice thicknesses compared with manual ECG-gated calcium-scoring CT (CSCT). Methods This retrospective study included 567 patients who underwent both LDCT and CSCT. All LDCT images were reconstructed with a 2.5-mm slice thickness (LDCT2.5-mm), and 453 LDCT scans were reconstructed with a 1.0-mm slice thickness (LDCT1.0-mm). Automated CAC scoring was performed on CSCT (CSCTauto), LDCT1.0-mm, and LDCT2.5-mm images. The reliability of CSCTauto, LDCT1.0-mm, and LDCT2.5-mm was compared with manual CSCT scoring (CSCTmanual) using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. Agreement, in CAC severity category, was analyzed using weighted kappa statistics. Diagnostic performance at various Agatston score cutoffs was also calculated. Results CSCTauto, LDCT1.0-mm, and LDCT2.5-mm demonstrated excellent agreement with CSCTmanual (ICC [95% confidence interval, CI]: 1.000 [1.000, 1.000], 0.937 [0.917, 0.952], and 0.955 [0.946, 0.963], respectively). The mean difference with 95% limits of agreement was lower with LDCT1.0-mm than with LDCT2.5-mm (19.94 [95% CI, −244.0, 283.9] vs. 45.26 [−248.2, 338.7]). Regarding CAC severity, LDCT1.0-mm achieved almost perfect agreement, and LDCT2.5-mm achieved substantial agreement (kappa [95% CI]: 0.809 [0.776, 0.838], 0.776 [0.740, 0.809], respectively). Diagnostic performance for detecting Agatston score ≥ 400 was also higher with LDCT1.0-mm than with LDCT2.5-mm (F1 score, 0.929 vs. 0.855). Conclusions Fully automated CAC-scoring software with both CSCT and LDCT yielded excellent reliability and agreement with CSCTmanual. LDCT1.0-mm yielded more accurate Agatston scoring than LDCT2.5-mm using fully automated commercial software.

키워드

Artificial intelligence; Software; Coronary arteries; Calcium; Tomography; X-ray computed; CORONARY-ARTERY CALCIUM; ELECTRON-BEAM CT; COMPUTED-TOMOGRAPHY; CALCIFICATION; ASSOCIATION; VALIDATION; AGREEMENT
제목
Evaluation of fully automated commercial software for Agatston calcium scoring on non-ECG-gated low-dose chest CT with different slice thickness
저자
Kang, Hyun Woo; Ahn, Woo Jin; Jeong, Ju Hyun; Suh, Young Joo; Yang, Dong Hyun; Choi, Hangseok; Hwang, Sung Ho; Yong, Hwan Seok; Oh, Yu-Whan; Kang, Eun-Young; Kim, Cherry
DOI
10.1007/s00330-022-09143-1
발행일
2023-03
유형
Article; Early Access
저널명
European Radiology
권
33
호
3