상세 보기
초록
The effect of fresh gas flow (FGF) on isoflurane concentrations at given vaporizer settings during low-flow anaesthesia was investigated. Ninety patients (American Society of Anaesthesiologists physical status I or II) were randomly allocated to three groups (FGF 1 l/min, FGF 2 l/min and FGF 4 l/min). Anaesthesia was maintained for 10 min with vaporizer setting isoflurane 2 vol% and FGF 4 l/min for full-tissue anaesthetic uptake in a semi-closed circle system. Low-flow anaesthesia was maintained for 20 min with end-tidal isoflurane 1.5 vol% and FGF 2 l/min. FGF was then changed to FGF 1 l/min, FGF 2 l/min or FGF 4 l/min. Measurements during the 20-min period showed that inspired and end-tidal isoflurane concentrations decreased in the FGF 1-l/min group but increased in the FGF 4-l/min group compared with baseline values. No hoemodynamic changes were observed. Monitoring of anaesthetic concentrations and appropriate control of vaporizer settings are necessary during low-flow anaesthesia.
키워드
- 제목
- Effect of fresh gas flow on isoflurane concentrations during low-flow anaesthesia
- 저자
- Park, JY; Kim, JH; Kim, WY; Chang, MS; Kim, JY; Shin, HW
- 발행일
- 2005-09
- 유형
- Article
- 권
- 33
- 호
- 5
- 페이지
- 513 ~ 519
- 언어
- ENG
- 출판사
- Cambridge Medical Publications
- 발행국가
- 영국
- 분량
- 7 페이지
- ISSN
- E 1473-2300
P 0300-0605