Assay of omeprazole and omeprazole sulfone by semi-microcolumn liquid chromatography with mixed-function precolumn

Citations

WEB OF SCIENCE

17
Citations

SCOPUS

20

초록

A column-switching system based on semi-microcolumns was used for direct analysis of omeprazole and omeprazole sulfone in human plasma samples. Plasma samples were injected into a mixed-function (MF Ph-1) column (35 mm×4.6 mm I.D.) to remove proteins and other non-specific peak producing substances from the analyte-containing time zone. The analyte-containing fraction was thereafter transferred to a C-18 semi-microcolumn (250 mm×1.5 mm I.D.) after concentration at the C-18 intermediate column. The absorbance at 302 nm in a ultraviolet (UV) detector was recorded to measure the concentration. The detection limit for omeprazole and omeprazole sulfone in the present method was 10 ng/ml. Interbatch variation (coefficient of variation) of the QC samples spanned less than 10% and intra-batch variation less than 2%. The recovery ratios of omeprazole and omeprazole sulfone were over 98%. The current method can be used as a simpler procedure with similar sensitivity and reproducibility as previously reported methods.

키워드

Column switching; Omeprazole; Omeprazole sulfone; ACID-SECRETION; METABOLITES; PLASMA; PHARMACOLOGY; INHIBITION; URINE
제목
Assay of omeprazole and omeprazole sulfone by semi-microcolumn liquid chromatography with mixed-function precolumn
저자
임동석; 정주은; 박지영
DOI
10.1016/S0378-4347(01)00036-6
발행일
2001-04
저널명
Journal of Chromatography - Biomedical Applications
권
754
호
2
페이지
487 ~ 493