A high-resolution lensless fluorescence imaging system using membrane deflection for reducing gap between samples and image sensor
- Shin, Hyogeun;
- Choi, Woongsun;
- Yoon, Gun-Wook;
- Seo, Suyoung;
- Choi, Nakwon;
- ... Cho, Il-Joo;
- 외 1명
WEB OF SCIENCE
0SCOPUS
0초록
This paper reports on a high-resolution lensless fluorescence imaging system using membrane deflection for reducing the gap between fluorescent samples and a CMOS image sensor array. The presented system applied the principle of the total internal reflection of excitation light inside a PDMS chip to minimize the exposure of excitation light to the image sensor. We integrated a very thin (5.5 mu m) PDMS membrane as a flexible deflector, which allowed for significant reduction of the gap (-20 mu m) inevitably created during the conventional packaging (i.e., wire bonding) of image sensors. In the fabricated system, we achieved minimum distance of 7.5 mu m between fluorescent samples and an image sensor array. This reduced gap almost directly corresponds to a resolution as high as 7.5 mu m, based on the point spread function (PSF).
- 제목
- A high-resolution lensless fluorescence imaging system using membrane deflection for reducing gap between samples and image sensor
- 저자
- Shin, Hyogeun; Choi, Woongsun; Yoon, Gun-Wook; Seo, Suyoung; Choi, Nakwon; Yoon, Jun-Bo; Cho, Il-Joo
- 발행일
- 2018-04
- 유형
- Proceedings Paper
- 저널명
- Proceedings of the IEEE International Conference on Micro Electro Mechanical Systems (MEMS)
- 페이지
- 419 ~ 421
- 언어
- ENG
- 출판사
- IEEE
- 분량
- 3 페이지
- ISSN
- P 1084-6999