Comparative study of i-SENS glucometers in neonates using capillary blood samples
WEB OF SCIENCE
3SCOPUS
3초록
The accuracy of point-of-care blood glucometers in the detection and evaluation of neonatal hypoglycemia is a concern. This study compared the performance of three i-SENS glucometers with that of the YSI 2300 STAT Plus Analyzer, which was used as a reference. The leftover neonatal capillary blood samples of 319 patients were used in this study. The evaluation process and accuracy performance criteria were based on the International Organization for Standardization 15197:2013 guidelines. The evaluation involved three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK® Inform II glucometer. The accuracy evaluation yielded acceptable results as follows: a) 100 and 100% for BAROzen H Expert plus; 99.8 and 100% for CareSens PRO; 98.7%, and 97.2% for CareSens H Beat glucometers were within the range of ±0.8 mmol/L (15 mg/dL) and ±15% of the average reference values at glucose concentrations <5.55 mmol/L (100 mg/dL) and ≥5.55 mmol/L (100 mg/dL), respectively, and b) all estimated glucose values (100%) were within the zones A and B of Consensus Error Grid for all three i-SENS glucometers. There was good correlation between the glucose values estimated by the glucometers and the reference values (R>0.990). This study demonstrated that i-SENS glucometers exhibit acceptable performance and can be used as effective point-of-care devices in neonates. © 2021 Walter de Gruyter GmbH, Berlin/Boston.
키워드
- 제목
- Comparative study of i-SENS glucometers in neonates using capillary blood samples
- 저자
- Kim, Ha Nui; Yoon, Soo-Young
- 발행일
- 2021-05
- 유형
- Article in Press
- 권
- 59
- 호
- 6
- 페이지
- 1133 ~ 1141
- 언어
- ENG
- 출판사
- Walter de Gruyter GmbH
- 발행국가
- 독일
- 분량
- 9 페이지
- ISSN
- E 1437-4331
P 1434-6621