Early malfunction of a biliary self-expandable metal stent with an antireflux valve: A case report

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Sang Hoon-
dc.contributor.authorOh, Chi Hyuk-
dc.contributor.authorLee, Jae Min-
dc.contributor.authorChoi, Seong Ji-
dc.contributor.authorChoi, Hyuk Soon-
dc.contributor.authorKim, Eun Sun-
dc.contributor.authorKeum, Bora-
dc.contributor.authorJeen, Yoon Tae-
dc.contributor.authorChun, Hoon Jai-
dc.contributor.authorLee, Hong Sik-
dc.contributor.authorKim, Chang Duck-
dc.date.available2020-12-09T08:48:53Z-
dc.date.issued2020-04-
dc.identifier.issn0025-7974-
dc.identifier.issn1536-5964-
dc.identifier.urihttps://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/33758-
dc.description.abstractRationale: Biliary drainage is essential to resolve jaundice in patients with malignant biliary obstruction. Recently, a biliary self-expandable metal stent (SEMS) with an antireflux valve was developed to prevent enteric-biliary reflux. Its antireflux valve was made of expanded polytetrafluoroethylene (ePTFE), a biostable and biocompatible material. Changes of the ePTFE membrane of medical devices are rarely reported in clinical practice. Patient concerns: A 59-year-old woman was admitted with a complaint of jaundice. Around 1 month before, she underwent endoscopic biliary stenting using a SEMS with an antireflux valve. Diagnosis: Acute cholangitis due to stent clogging was detected on abdominal computed tomography. Interventions: Endoscopic retrograde cholangiopancreatography and removal of the SEMS with deformity were performed. Outcomes: The obstructive jaundice was rapidly resolved after the removal of the clogged stent. Lessons: Although the ePTFE membrane is widely used to make medical devices, unexpected changes could occur after the membrane is exposed to bile juice in the duodenum. When performing biliary stenting using a SEMS with an antireflux valve, both the risk of early malfunction and the benefit of the antireflux effect should be considered.-
dc.language영어-
dc.language.isoENG-
dc.publisherLippincott Williams & Wilkins Ltd.-
dc.titleEarly malfunction of a biliary self-expandable metal stent with an antireflux valve: A case report-
dc.typeArticle-
dc.publisher.location미국-
dc.identifier.doi10.1097/MD.0000000000019750-
dc.identifier.scopusid2-s2.0-85084627401-
dc.identifier.wosid000544714800057-
dc.identifier.bibliographicCitationMedicine, v.99, no.16-
dc.citation.titleMedicine-
dc.citation.volume99-
dc.citation.number16-
dc.type.docTypeArticle-
dc.description.isOpenAccessN-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaGeneral & Internal Medicine-
dc.relation.journalWebOfScienceCategoryMedicine, General & Internal-
dc.subject.keywordPlusOBSTRUCTION-
dc.subject.keywordPlusPALLIATION-
dc.subject.keywordPlusMULTICENTER-
dc.subject.keywordPlusSTRICTURES-
dc.subject.keywordPlusREDUCE-
dc.subject.keywordAuthorantireflux valve-
dc.subject.keywordAuthorbile duct obstruction-
dc.subject.keywordAuthorexpanded polytetrafluoroethylene-
dc.subject.keywordAuthormalfunction-
dc.subject.keywordAuthorself-expandable metal stent-
Files in This Item
There are no files associated with this item.
Appears in
Collections
5. Others > Others(Medicine) > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Hyuk Soon photo

Choi, Hyuk Soon
Anam Hospital (Department of Gastroenterology and Hepatology, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE