Measurement of Hemodynamic Energy at Different Vessels in an Adult Swine Model

Authors
Son, Kuk HuiAhn, Chi BumLee, Sung HoSon, Ho SungChoi, JeasoonJung, Jae SeungSun, KyungKim, Kwang Taik
Issue Date
Sep-2010
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Citation
ASAIO JOURNAL, v.56, no.5, pp 397 - 402
Pages
6
Indexed
SCI
SCIE
SCOPUS
Journal Title
ASAIO JOURNAL
Volume
56
Number
5
Start Page
397
End Page
402
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2020.sw.kumedicine/14591
DOI
10.1097/MAT.0b013e3181e6c2c7
ISSN
1058-2916
1538-943X
Abstract
Extra hemodynamic energy is one of the major benefits of pulsatile flow, improving blood flow to vital organs. But most (80%) of the hemodynamic energy generated from pulsatile flow is damped by the extracorporeal circuit. Most models devised to minimize hemodynamic energy loss have been in vitro pediatric models. The purpose of this study was to measure hemodynamic energy in different vessels of different organs with an in vivo adult swine model. An extracorporeal circuit was constructed for seven Yorkshire swine using a pulsatile pump (Twin-Pulse Life Support). The mean arterial pressure (MAP), energy equivalent pressure (EEP), and surplus hemodynamic energy (SHE) at the renal artery, carotid artery, aortic cannula site, and postoxygenator site were measured simultaneously before starting the pump and at the pump rates of 25, 35, and 45 bpm. The MAP of the renal or carotid artery was 40.0%-51.2% of the postoxygenator site. The EEP and SHE of both arteries were 11.6%-13.0% and 5.5%-7.4% of the postoxygenator site, respectively. The MAP and EEP of both arteries after starting the pump were lower than at baseline. The SHE of the renal artery after starting the pump was significantly higher than at baseline. The SHE of the carotid artery increased substantially after starting the pump although not statistically significantly. There was a significant hemodynamic energy loss in both arterial sites compared with the postoxygenator site. Also, a difference in hemodynamic energy loss was observed in vessel-to-vessel or vessel-to-circuit site comparison. This difference creates a bias in studying pulsatility and its effects. Therefore, the measurement method of hemodynamic energy must be standardized and the measurement site clarified to yield accurate study results. ASAIO Journal 2010; 56:397-402.
Files in This Item
There are no files associated with this item.
Appears in
Collections
2. Clinical Science > Department of Thoracic and Cardiovascular Surgery > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Son, Ho Sung photo

Son, Ho Sung
Anam Hospital (Department of Thoracic and Cardiovascular Surgery, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE