Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Performance and Physician Experience of INGEVITY plus Active Fixation Leads: Prospective INGEVITY plus Lead Clinical Study in Koreaopen access

Authors
Joung, BoyoungBae, Myung HwanOh, Il-YoungPark, Hyung-SeobShim, JaeminCho, Min SooLee, Jung MyungChoi, Eue-KeunLee, Young Soo
Issue Date
Jan-2024
Publisher
Hindawi Publishing Corporation
Citation
Cardiology Research and Practice, v.2024
Indexed
SCIE
Journal Title
Cardiology Research and Practice
Volume
2024
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2021.sw.kumedicine/65358
DOI
10.1155/2024/2172306
ISSN
2090-8016
2090-0597
Abstract
Background. Boston Scientific INGEVITY+ pacing lead (Boston Scientific, Marlborough, MA, USA) has been upgraded to INGEVITY. The performance of the INGEVITY+ pacing lead has not yet been reported. This study aimed to evaluate the short- and long-term safety, effectiveness, and handling experience of INGEVITY+ leads. Methods. Consecutive patients were included from 9 institutions in Korea, where 400 leads (200 right ventricular active fixation leads and 200 right atrial active fixation leads) were implanted or attempted in 200 subjects. Results. During the implantation, only one patient required a lead change because of lead screw failure. The handling questionnaires of the lead received very positive feedback with 88% of operators agreeing that it is easy for leads to pass through small vessels or vessels with multiple leads. At the 3-month follow-up, 95.7% of RA leads and 99.5% of RV leads had pacing thresholds less than 1.5 V. A total of 92.4% of atrial leads had amplitudes greater than 1.5 mV, and 96.5% of ventricular leads had sensing amplitudes greater than 5 mV at 3 months. A total of 99.8% had impedances between 300 and 1,300 ohms. The lead-related complication-free rate for all leads during follow-up was 100%, and the overall rates of lead dislodgment, perforation, and pericardial effusion were all 0.0%. Conclusions. The INGEVITY+ pacing lead exhibited exceptional clinical performance, with a high complication-free rate throughout the 3-month follow-up period. In addition, the lead displayed excellent electrical characteristics, and the lead-handling experience was reported to be very good.
Files in This Item
There are no files associated with this item.
Appears in
Collections
5. Others > ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Shim, Jae min photo

Shim, Jae min
Anam Hospital (Department of Cardiology, Anam Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE