Comparison of retinal thickness measurements among four different optical coherence tomography devicesopen access

Authors
Nam, Ki TaeYun, CheolminSeo, MyunghoAhn, SominOh, Jaeryung
Issue Date
Feb-2024
Publisher
Nature Publishing Group
Citation
Scientific Reports, v.14, no.1
Indexed
SCIE
SCOPUS
Journal Title
Scientific Reports
Volume
14
Number
1
URI
https://scholarworks.korea.ac.kr/kumedicine/handle/2021.sw.kumedicine/65760
DOI
10.1038/s41598-024-54109-6
ISSN
2045-2322
Abstract
We sought to compare the retinal thickness measurements collected using different optical coherence tomography (OCT) devices. This prospective study included 21 healthy cases, and the retinal thickness was measured using the PLEX Elite (Carl Zeiss Meditec, Dublin, California, USA), DRI OCT-1 Atlantis (Topcon Corp, Tokyo, Japan), Cirrus 5000 HD-OCT (Carl Zeiss Meditec), and Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany), respectively. The mean central retinal thickness (CRT) and mean retinal thickness of the Early Treatment of Diabetic Retinopathy Study (ETDRS) area were compared. The CRT varied significantly among the different OCT devices (P < 0.001). Post-hoc analysis revealed that the CRT measured using PLEX Elite (278.95 +/- 20.04 <mu>m) and Spectralis (271.86 +/- 17.92 mu m) were similar, and both were greater than the CRT measurements of DRI OCT-1 (239.57 +/- 21.06 mu m) and Cirrus (256.76 +/- 17.82 mu m). Additionally, the mean retinal thickness in each ETDRS area showed significant differences among the four devices (all P < 0.001). The mean retinal thickness measured varied according to the device used, and this needs to be considered when comparing retinal thickness measurements taken with different devices.
Files in This Item
There are no files associated with this item.
Appears in
Collections
5. Others > ETC > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yun, Cheol min photo

Yun, Cheol min
Ansan Hospital (Department of Ophthalmology, Ansan Hospital)
Read more

Altmetrics

Total Views & Downloads

BROWSE